Selecting the Right Pumping System for Your Device
by Eric Pepe, KNF Neuberger, Inc.
Its the kind of scenario that makes engineers and medical device designers cringe. Youve finished designing a $50,000 blood analyzer and, with 50 units in the field, youre already moving on to your next project...
Then, the news comes in. Because of an inappropriate pump in the system, virtually all of the new analyzers have failed to function properly. Critical development time and moneyas well as tightly knit plans for marketing the new producthave suffered a major setback.
This is not an imaginary scenario. Nor is it a rare one. While in the past 10 years major technological breakthroughs have occurred in the design of pumping systemsgiving design engineers more flexibility than ever beforemany engineers are not aware of them. Additionally, clear guidelines highlighting important criteria for selecting the right pump have been seriously lacking. Add to this mix the medical device designers traditional approach to purchasing a pumplate in the design process and from a catalog of standard productsand you have a recipe for disaster.
To avoid a pumps failing, it is extremely important that the designer acquaint himself with the latest pumping system technology and clarify the key parameters of his pumping systems needs. Then, and most important, the designer must communicate these parameters to his pump supplier early in the design process.
Pumping Systems Defined
A pump is a subsystem, not a commodity. It is a dynamic, interactive element of the medical device in which it functions. A pumping system will perform differently as conditions within the medical device change. For example, temperature or electrical-power variations beyond a pumps defined tolerance limits could cause the pump to malfunction and thus the entire medical system to shut down.
Pump Types
There are many types of pumping systems on the market, each offering iis own advantages for particular applications. Examples include:
Critical Performance Requirements
A pumps performance can vary dramatically depending on the device and the environment in which it operates. Therefore, it is vital to clarify early in the design process the systems performance requirements and to match them with those of the devices design. Several critical performance requirements to consider are:
It is critical that designers specify the pumping systems tolerance to various performance requirements, including electrical power, temperature, flow rate, vacuum, and pressure. Specifying a pumps power requirement, for example, without toleranceat 115 Vis not enough. Rather, if the pump operates in a system that varies up to 10 V, specify 115 V with a tolerance of +- 10 V. Similarly, when specifying vacuum rate and flow rate, the designer needs to specify a tolerance of flow rate over a tolerance of vacuum or pressure. This helps ensure that the pumping system will function in the particular system environment for which it was designed.
For example, if a pump has the tolerance to create a vacuum greater than that required by the device and the device contains soft tubing, the pumps excessive vacuum could cause the soft tubing, the pumps excessive vacuum could cause the soft tubing to collapse, resulting in a system shut-down or equipment damage. Likewise, a pumping system that can create pressure greater than a medical instruments tolerance can break connectors and other system parts when the pressure becomes excessive.
Lets look at some of these requirements in more detail.
Electrical Requirement
Naturally, every system environment demands specific electrical-power requirements. Will the pumping system operate on 115 V/60 cycles exclusively? Will the system operate in the United States as well as Europe, requiring that it operate with both 50-and 60-cycle current?
Perhaps voltage conditioning will be a factorrequiring a voltage transformer between the pump and power supplyso that the pump operates with 115 V in one environment and 220 V in another.
If a pump designated to operate at 115 V (+=10 V) is located where voltage may vary as high as 130 V, that pump may shut down, resulting in overall system failure.
Ambient Temperature
The pumps location within an instrument is very important. If ventilation is in adequate, the ambient temperature could climb. If it climbs above the temperature tolerance of the pump, the system could shut down and the equipment fail. At the very least, high ambient temperature and improper ventilation can shorten the life span of a pumping system.
Duty Cycle
Some blood analyzers, for example, will at some point in their operation perform an analytical function, such as optically reading the contents of a vial. This function can take from 10 seconds to two minutes. During this time, fluid movement within the system stops; the vacuum pump is not needed and frequently turns off. How long the pumping system remains inactive, the demands on the pump when the analyzer performs its next function, and many other factors can be critical to selecting the right pump for the job.
For instance, will the pump need to restart against load? Most pumping systems must be at no-load conditions on inlet and outlet to restart properly. However, some pumps can be modified to start against vacuum. If a pumping system must restart against vacuum, the designer must communicate this to the pump supplier. Otherwise, a standard pumpone not designed to start against pressure may be supplied.
Case Histories
The following case histories illustrate some common pitfalls designers encounter when selecting a pumping system.
Case 1: A designer selected a linear piston pump to perform a vacuum application in a blood analyzer that handled wet gases (an inappropriate application for this type of pump). Of the 50 units in the field, virtually all were failing. The linear pumps were not producing the necessary vacuum. Moisture was corroding the piston and cylinder, rendering the pumps inoperable. The solution: The piston pump was removed, the architecture of the vacuum system was modified, and three small diaphragm liquid vacuum pumps were incorporated.
Case 2: A sterilizer manufacture had improperly defined the operating parameters of his unit for maximum/minimum pressure. The manufacturer selected a pump with a maximum pressure of 43 psi. While this was the pumps maximum pressure, it was also the minimum pressure needed for the sterilizer to function. Therefore, when the pumps pressure dropped below 43 psi, the sterilizer did not function. This was the case with units operating in cities where absolute pressure is lower because of the altitude. Solution: the operating parameters for pressure were properly defined and the pump was modified.
Case 3: The designer of a blood analyzer specified a pump with a 115-V, 50-and 60-cycle, motor power requirement. However, he did not specify a voltage power tolerance. The analyzers were operated in England where voltage output is 10% higher than in the rest of Europe. The pumps motor coil received approximately 127 V at 50 cycles, which was beyond the motors capability. The results were thermal overload and pump shut-off. Solution: a larger motor with a specially wound coil was incorporated into the pumping system. The pump now has a voltage tolerance of at least 10%.
ConclusionPump failures, such as the ones illustrated above, usually are not the fault of the pumping system, but rather of the designers not communicating complete system requirements to the pump supplier. Working closely together, designers and suppliers can select the right pump to match a systems requirements.
Entire Web Site Copyright (c) 2010 by KNF Neuberger, Inc. (USA) - All Rights Reserved.