Return to KNF USA Home Page
Call us at 609-890-8600
To KNF USA Home Page
Micro PumpsMini PumpsHigh CapacitySwing PistonHeated SamplingDouble DiaphraExplosion ProofDeep Vacuum
Request a Quote or Evaluation UnitDownload KNF's OEM BrochureFrequently Asked QuestionsA collection of technical articles on diaphragm pumpsPress Releases and New Products Information
 
Heated Head/Heat Resistant OEM Pumps Search:
Pump Series
(Link to Datasheets)
Available
Motors
Flow Rate
LPM
Flow Rate
CFM
Max Vacuum
in. Hg
Max Pressure
psig
N86 (PDF:925kb/4pgs) NEW!

AC

6 .21 21 22
N012  (PDF:925kb/4pgs) NEW! AC 10.5 .37 22.8 22
N024  (PDF:925kb/4pgs) NEW! AC 18 .64 24 22
N036  (PDF:925kb/4pgs) NEW! AC 34 1.2 24 20
N036.0  (PDF:225kb/4pgs) NEW! AC 30 per head 1 per head 24 20
Motor Key: AC=Alternating Current.

The model N024ST.26E stainless steel heated-head sampling pump with included insulating jacket for thermal stability and safety, is available with electronic temperature controller and RS-232 interface.

The model N86ST.16I mini heat-resistant sampling pump is available in aluminum or 316 stainless steel for analytical instruments. It is excellent for use in smaller analyzers.

These pumps are popular for use in exhaust gas and stack gas sampling systems.

About Heated-Head Pumps and Controls

Equipped for the efficient transport of hot gas samples in environmental and analytical instruments, they handle gas temperatures up to 460°F (240°C). The bearings and motor are isolated from the heat for longer component life. The heated-head model uses either an electrically-heated cartridge regulated by a thermostat, or an optional programmable temperature controller (see example at right), to maintain specified head temperature. The efficient insulation jacket (not N86) stabilizes head temperature, reduces heat loss and prevents accidental contact.
Models are available as:

  • (.16) Heat-resistant model
  • (.11) Heated-head model temperature regulated by mechanical thermostat
  • (.26) Heated-head model temperature regulated by programmable controller


Control panel of programmable controller

Unit Conversion Tools:

Your browser's Javascript function must be turned on to use the Unit Conversion Tools.

NOTES: Standard continuous performance ratings listed on this KNF website and on KNF datasheets are based on technical data and test results of nominal units at sea level, measured with an ambient temperature of 70°F (21°C), and nominal electrical supply as listed above, and are for reference only.Multiple-head models, different performance models, cost-effective OEM modifications, and a variety of motor options and different materials of construction are also available. Certified drawings are available upon request. Specifications subject to change without notice. You should always check corrosion resistance charts for compatibility with the medium you intend to pump.

KNF NEUBERGER, INC.
Two Black Forest Road
Trenton, New Jersey 08691-1810
609-890-8600 • Fax 609-890-8323 • site map